Key Product Information


Specifications Select a tab below for more information



Detection: Isotropic (X, Y and Z Axis Readings) 
Frequency Range: 100 kHz to 6 GHz 
Frequency Response with Correction: 100 kHz to 6 GHz ± 0.9 dB
Frequency Response: (Typical)
  • 500 kHz to 2 GHz +1.0, -2.5 dB
  • 2 GHz to 5.5 GHz +3.5, -4.5 dB
  • 5.5 GHz to 6 GHz +2.0, -6.0 dB
Dynamic Range:  0.5 V/m to 800 V/m (Single Range) 
Resolution: 0.01 V/m 
Isotropic Deviation: ±0.5 dB @ 400 MHz 
Linearity: ±0.5 dB (1 to 800 V/m)   
Sample Rate (Typical): 70 Samples per Second Maximum   
Overload Withstand: >1500 V/m CW
Physical Interface:
  • Duplex Optical Fiber (62.5 Micron Multimode)
  • FC Connectors for Laser Cable, Integral 1m Optical Cable
  • ST Connector for Transmitter Cable, Integral 1m Optical Cable
Operating Range: 10°C to 40°C (50° F to 104° F); 5 to 95% Relative Humidity Non-condensing 
Dimensions (Housing): 32 mm x 32 mm x 32 mm; (1.26 in x 1.26 in x 1.26 in) 
Mounting:  1/4 in - 20 UNC Internal Thread         
  • Manual
  • Probe Assembly
  • A2LA Traceable Calibration Report
  • 10m Optical Cable
  • Bulkhead Connector (2)
  • Carrying Case
  • Fiber Cleaning Kit and Swabs

EMCenter Interface Card (125428) 

Fiber Optic Maintenance Kit (112333)

Probe Stand (H-491269)

Extra Probe Carriers for above Probe Stand (H-491276)

HI-6100 Field Monitor Interface Card (H-491277) 

 

 

 

Key Features


  • Laser Powered -Permits Extended Testing
  • Frequency Response: 
    • 100 kHz to 6 GHz
    • Dynamic Range 0.5 V/m to 800 V/m (Single Range) 
  • Provides Individual and Summed Axis Values
  • A2LA Accredited Calibration Report
  • Suitable for MIL Standard Specs:
    •  MIL-STD 461F Radiated Susceptibility (RS) 
  • Suitable for Automotive Specs:
    • SAE J1113 27
    • GMW 3091/3097/3103
    • FORD FMC 1278
  • Suitable for Commercial Specs: EN/IEC 61000-4-3 Radiated Immunities
  • Operates with Most 3rd Party Immunity Software

 For High Field and High Frequency Requirements the Model HI-6x53 is Recommended