Product Image

Key Features

  • Provides an Interface for ETS-Lindgren Battery-Powered Electrical Field Probes
  • Wide Frequency and Dynamic Ranges
  • Compact Design
  • Download PDF Datasheet
  • ETS-Lindgren EMSense EMF Probe Plug-in Card provides an interface for ETS-Lindgrens battery-powered electrical field probes to the EMCenter Modular RF Platform.

    Each EMSense Modular Plug-in Card supports one ETS-Lindgren electrical field probe. EMSense and ETS-Lindgren EMC field probes are fully supported by TILE and other test automation software packages. Please contact ETS-Lindgren for additional information.

    Specifications
    View the EMSense EMF Probe Plug-in Card Model 7007-001 technical specifications below

    Physical Specifications

    Specification
    7007-001 (Plug-in Card) HI-6023 (Optional E-field Probe) HI-6006 (Optional E-field Probe) HI-6053 (Optional E-field Probe)
    Exterior Dimension
    1 Slot
    32 mm x 32 mm Housing
    (1.25 in x 1.25 in x 1.25 in)
    32 mm x 32 mm Housing
    (1.25 in x 1.25 in x 1.25 in)

    76 mm x 51 mm x 51 mm

    Housing (3.0 in x 2.0 in x 2.0 in)


    Sensor Protective Caps Height
    n/a
    36 mm (1.42 in)
    43 mm (1.69 in)
    348 mm x 58.4 mm (13.7 in probe length x 2.3 in width)
    Weight
    n/a
    80 g (2.82 oz)
    80 g (2.82 oz)
    360 g (12.7 oz)
    Mounting
    n/a
    1/4 20 - UC Internal Thread
    1/4 20 - UC Internal Thread
    1/4 20 - UC Internal Thread
    Physical Interface
    n/a
    Duplex Optical Fiber
    (200 Micron Multi-mode)
    FSMA Connectors
    Integral 1m Optical Cable

    Duplex Optical Fiber
    (200 Micron Multi-mode)
    FSMA Connectors
    Integral 1m Optical Cable
    Duplex Optical Fiber
    (200 Micron Multi-mode)
    FSMA Connectors
    Temperature Range
    0 C to +35 C
    (32F to +95F)

    +10 to +40 C
    (+50F to +95F)

    +10 to +40 C
    (+50F to +95F)

    +10 to +40 C
    (+50F to +95F)

    Relative Humidity
    10 to 90% (Non-condensing)
    5 to 95% (Non-condensing)
    5 to 95% (Non-condensing)
    5 to 95% (Non-condensing)
    Battery (Field replaceable)
    n/aRechargeable NiMH
    Rechargeable NiMHRechargeable NiMH AAA 900mA/hr
    Battery Life
    n/a8 Hours8 Hours30 Hours
    Battery Charger
    n/a100 - 240 VAC Universal Input
    3-Hour Charge from Full Depletion
    100 - 240 VAC Universal Input
    3-Hour Charge from Full Depletion
    100 - 240 VAC Universal Input
    3-Hour Charge from Full Depletion

    Electrical Specifications

    Specification
    HI-6023 (Optional E-field Probe)HI-6006
    (Optional E-field Probe)

    HI-6053 (Optional E-field Probe)
    Detection
    Isotropic (X, Y and Z Axis Readings)
    Isotropic (X, Y and Z Axis Readings)
    Isotropic (X, Y and Z Axis Readings)
    Calibrated Frequency Range
    10 kHz to 1 GHz
    100 kHz to 6 GHz
    10 MHz to 40 GHz
    Frequency Response (Typical)
    10 kHz to 30 kHz +0.5, - 2.5 dB
    30 kHz to 1 GHz 1.0 dB



    500 kHz to 2 GHz +1.0, -2.5 dB
    2 to 5 GHz +3.5, -4.0 dB
    5 to 6 GHz +1.0, -6.0 dB

    10 to 100 MHz + 3.0, -4.0 dB
    100 MHz to 1 GHz +3.0, -0.50 dB
    1 to 18 GHz +4.0, -2.0 dB
    18 to 40 GHz +3.0, -4.5 dB
    Frequency Response with Correction
    0.9 dB
    0.9 dB
    10 MHz to 18 GHz 0.9 dB
    18 to 40 GHz 1.1 dB
    Dynamic Range
    2.0 to 800 V/m (Single Range)
    0.5 to 800 V/m (Single Range)
    2.0 to 800 V/m (Single Range)
    Resolution
    0.01 V/m
    0.01 V/m
    0.01 V/m
    Isotropic Deviation
    0.5 dB @ 400 MHz
    0.5 dB @ 400 MHz
    1.0 dB <18 GHz
    Linearity
    0.5 dB @ 27 MHz
    0.5 dB @ 27 MHz
    0.5 dB @ 1 GHz
    Sample Rate
    60 Samples/Second Maximum
    60 Samples/Second Maximum
    60 Samples/Second Maximum
    Overload Withstand
    >1500 V/m CW
    >1500 V/m CW
    >1500 V/m CW

    Product Configuration

    Specification
    HI-6023 (Optional E-field Probe)HI-6006
    (Optional E-field Probe)

    HI-6053 (Optional E-field Probe)
    Detection
    Isotropic (X, Y and Z Axis Readings)
    Isotropic (X, Y and Z Axis Readings)
    Isotropic (X, Y and Z Axis Readings)
    Calibrated Frequency Range
    10 kHz to 1 GHz
    100 kHz to 6 GHz
    10 MHz to 40 GHz
    Frequency Response (Typical)
    10 kHz to 30 kHz +0.5, - 2.5 dB
    30 kHz to 1 GHz 1.0 dB



    500 kHz to 2 GHz +1.0, -2.5 dB
    2 to 5 GHz +3.5, -4.0 dB
    5 to 6 GHz +1.0, -6.0 dB

    10 to 100 MHz + 3.0, -4.0 dB
    100 MHz to 1 GHz +3.0, -0.50 dB
    1 to 18 GHz +4.0, -2.0 dB
    18 to 40 GHz +3.0, -4.5 dB
    Frequency Response with Correction
    0.9 dB
    0.9 dB
    10 MHz to 18 GHz 0.9 dB
    18 to 40 GHz 1.1 dB
    Dynamic Range
    2.0 to 800 V/m (Single Range)
    0.5 to 800 V/m (Single Range)
    2.0 to 800 V/m (Single Range)
    Resolution
    0.01 V/m
    0.01 V/m
    0.01 V/m
    Isotropic Deviation
    0.5 dB @ 400 MHz
    0.5 dB @ 400 MHz
    1.0 dB <18 GHz
    Linearity
    0.5 dB @ 27 MHz
    0.5 dB @ 27 MHz
    0.5 dB @ 1 GHz
    Sample Rate
    60 Samples/Second Maximum
    60 Samples/Second Maximum
    60 Samples/Second Maximum
    Overload Withstand
    >1500 V/m CW
    >1500 V/m CW
    >1500 V/m CW