Model
7002-006 Plug-in Card Features
​Detector Type
​​​RMS Log Detector
​Measurement Function
​​​RMS CW Power, Peak (Max Hold) and Burst Mode​
Calibrated Frequency Range
​​10 MHz to 6 GHz​
​​Power Measuring Range
​​-50 dBm to +10 dBm
​​Maximum SWR​
​1.05 @ Below 1 GHz
1.15 @ 1 GHz to 6 GHz​
​Frequency Response Accuracy (at 23° C ± 2° C)​
​±0.25 dB​
​Deviation from CW for Wideband Modulated Signals wtih High Crest Factor
​​0.25 dB
​​Linearity Error​
​0.05 dB + 0.005 dB/dB
​​Input Damage Level
​> +20dB
​​Resolution​
​0.01 dB
​​RF Input Impedance
​​50 Ω​
​Measuring Speed
​​Max. 5 MS/s​
​Temperature Effect
​​0.15 dB Over Full Temperature Range
​​Zero Adjustment​
​Not Required
​​Frequency Response Correction​
​Stored Frequency Response Data is Taken into Account by Numerical Entry of the Measurement Frequency
​​Measurement Units
​​dBm/Watt

  • ​Interchangeable EUT Power Adapters: NEMA, Schuko, British or Australian Adapters Available

Key Features


  • Fully Compliant with ETSI 300 328 and 301 893
  • Fast, Accurate Performance
  • Easy Operation
  • Burst Mode