EMSenseā„¢ EMF Probe Plug-in Card Model 7007-001 - ETS-Lindgren
EMSenseā„¢ EMF Probe Plug-in Card Model 7007-001
Product Image
Key Features
  • Provides an Interface for ETS-Lindgren Battery-Powered Electrical Field Probes
  • Wide Frequency and Dynamic Ranges
  • Compact Design

ETS-Lindgren EMSense EMF Probe Plug-in Card provides an interface for ETS-Lindgrens battery-powered electrical field probes to the EMCenter Modular RF Platform.

Each EMSense Modular Plug-in Card supports one ETS-Lindgren electrical field probe. EMSense and ETS-Lindgren EMC field probes are fully supported by TILE and other test automation software packages. Please contact ETS-Lindgren for additional information.

Product Features

Wide Frequency and Dynamic Ranges

EMSense is fully compatible with all ETS-Lindgren EMC battery-powered electric field probes, covering a frequency range of 10 kHz to 40 GHz and a dynamic range of 0.5 to 800 V/m.

Compact Design

Each EMSense plug-in card occupies one slot in the EMCenter platform, allowing additional slots to be occupied by modular plug-in cards for positioning controllers, multiple RF switches and more.

Specifications
Physical Specifications
Specification
7007-001 (Plug-in Card) HI-6023 (Optional E-field Probe) HI-6006 (Optional E-field Probe) HI-6053 (Optional E-field Probe)
Exterior Dimension
1 Slot
32 mm x 32 mm Housing
(1.25 in x 1.25 in x 1.25 in)
32 mm x 32 mm Housing
(1.25 in x 1.25 in x 1.25 in)

76 mm x 51 mm x 51 mm

Housing (3.0 in x 2.0 in x 2.0 in)


Sensor Protective Caps Height
n/a
36 mm (1.42 in)
43 mm (1.69 in)
348 mm x 58.4 mm (13.7 in probe length x 2.3 in width)
Weight
n/a
80 g (2.82 oz)
80 g (2.82 oz)
360 g (12.7 oz)
Mounting
n/a
1/4 20 - UC Internal Thread
1/4 20 - UC Internal Thread
1/4 20 - UC Internal Thread
Physical Interface
n/a
Duplex Optical Fiber
(200 Micron Multi-mode)
FSMA Connectors
Integral 1m Optical Cable

Duplex Optical Fiber
(200 Micron Multi-mode)
FSMA Connectors
Integral 1m Optical Cable
Duplex Optical Fiber
(200 Micron Multi-mode)
FSMA Connectors
Temperature Range
0 C to +35 C
(32F to +95F)

+10 to +40 C
(+50F to +95F)

+10 to +40 C
(+50F to +95F)

+10 to +40 C
(+50F to +95F)

Relative Humidity
10 to 90% (Non-condensing)
5 to 95% (Non-condensing)
5 to 95% (Non-condensing)
5 to 95% (Non-condensing)
Battery (Field replaceable)
n/aRechargeable NiMH
Rechargeable NiMHRechargeable NiMH AAA 900mA/hr
Battery Life
n/a8 Hours8 Hours30 Hours
Battery Charger
n/a100 - 240 VAC Universal Input
3-Hour Charge from Full Depletion
100 - 240 VAC Universal Input
3-Hour Charge from Full Depletion
100 - 240 VAC Universal Input
3-Hour Charge from Full Depletion

Electrical Specifications
Specification
HI-6023 (Optional E-field Probe)HI-6006
(Optional E-field Probe)

HI-6053 (Optional E-field Probe)
Detection
Isotropic (X, Y and Z Axis Readings)
Isotropic (X, Y and Z Axis Readings)
Isotropic (X, Y and Z Axis Readings)
Calibrated Frequency Range
10 kHz to 1 GHz
100 kHz to 6 GHz
10 MHz to 40 GHz
Frequency Response (Typical)
10 kHz to 30 kHz +0.5, - 2.5 dB
30 kHz to 1 GHz 1.0 dB



500 kHz to 2 GHz +1.0, -2.5 dB
2 to 5 GHz +3.5, -4.0 dB
5 to 6 GHz +1.0, -6.0 dB

10 to 100 MHz + 3.0, -4.0 dB
100 MHz to 1 GHz +3.0, -0.50 dB
1 to 18 GHz +4.0, -2.0 dB
18 to 40 GHz +3.0, -4.5 dB
Frequency Response with Correction
0.9 dB
0.9 dB
10 MHz to 18 GHz 0.9 dB
18 to 40 GHz 1.1 dB
Dynamic Range
2.0 to 800 V/m (Single Range)
0.5 to 800 V/m (Single Range)
2.0 to 800 V/m (Single Range)
Resolution
0.01 V/m
0.01 V/m
0.01 V/m
Isotropic Deviation
0.5 dB @ 400 MHz
0.5 dB @ 400 MHz
1.0 dB <18 GHz
Linearity
0.5 dB @ 27 MHz
0.5 dB @ 27 MHz
0.5 dB @ 1 GHz
Sample Rate
60 Samples/Second Maximum
60 Samples/Second Maximum
60 Samples/Second Maximum
Overload Withstand
>1500 V/m CW
>1500 V/m CW
>1500 V/m CW

Product Configuration
Specification
HI-6023 (Optional E-field Probe)HI-6006
(Optional E-field Probe)

HI-6053 (Optional E-field Probe)
Detection
Isotropic (X, Y and Z Axis Readings)
Isotropic (X, Y and Z Axis Readings)
Isotropic (X, Y and Z Axis Readings)
Calibrated Frequency Range
10 kHz to 1 GHz
100 kHz to 6 GHz
10 MHz to 40 GHz
Frequency Response (Typical)
10 kHz to 30 kHz +0.5, - 2.5 dB
30 kHz to 1 GHz 1.0 dB



500 kHz to 2 GHz +1.0, -2.5 dB
2 to 5 GHz +3.5, -4.0 dB
5 to 6 GHz +1.0, -6.0 dB

10 to 100 MHz + 3.0, -4.0 dB
100 MHz to 1 GHz +3.0, -0.50 dB
1 to 18 GHz +4.0, -2.0 dB
18 to 40 GHz +3.0, -4.5 dB
Frequency Response with Correction
0.9 dB
0.9 dB
10 MHz to 18 GHz 0.9 dB
18 to 40 GHz 1.1 dB
Dynamic Range
2.0 to 800 V/m (Single Range)
0.5 to 800 V/m (Single Range)
2.0 to 800 V/m (Single Range)
Resolution
0.01 V/m
0.01 V/m
0.01 V/m
Isotropic Deviation
0.5 dB @ 400 MHz
0.5 dB @ 400 MHz
1.0 dB <18 GHz
Linearity
0.5 dB @ 27 MHz
0.5 dB @ 27 MHz
0.5 dB @ 1 GHz
Sample Rate
60 Samples/Second Maximum
60 Samples/Second Maximum
60 Samples/Second Maximum
Overload Withstand
>1500 V/m CW
>1500 V/m CW
>1500 V/m CW