Product Image

Key Features

  • Easy to Use
  • Accurate Measurements for EMC Test
  • Download PDF Datasheet
  • ETS-Lindgrens EMGen RF Signal Generator offers users a convenient and economical solution for signal generation. The EMGen Module is designed to fit into the EMCenter Modular Test System and provides the user with a signal generator that can be used in different modes across the 4 kHz to 6 GHz frequency range.

    Product Features
    View the EMGen RF Signal Generator Model 7003-003product features below

    Flexible Operation

    The EMGen EMC Signal Generator can be used in Continuous Wave (CW) mode, as well as AM, FM or pulse modulation mode.
    The EMGen covers a frequency range from 4 kHz to 6 GHz and can be used for both conducted and radiated immunity tests. It has a single SMA output connector.

    Accurate Measurements

    The EMGen is an accurate RF signal generator with a frequency error of +/- 1ppm/Year and a modulation accuracy of 1 Hz/< 1%. At the same time its output has an extremely fast response (less than 500s) which is very important for EMC testing.
    In addition, special engineering efforts have been taken to assure complete monotone behavior and prevent undesirable results at the output level. All these features make the EMGen the perfect signal generator for EMC immunity testing.

    Easy to Use

    The EMGen system is Plug-and-Play, meaning this card is automatically recognized and initialized by the EMCenter and is immediately ready to use. The user can configure and control the system by using TILE! Lab Management Software.

    Specifications
    View the EMGen RF Signal Generator Model 7003-003 technical specifications below

    Physical Specifications

    Dimensions: 3U x One 19 in Slot in EMCenter
    Temperature Range (Use): +10 C to +40 C
    Relative Humidity: 10 to 90% (Non-condensing)

    Electrical Specifications

    Model
    7003-003 Plug-in Card Feature
    Frequency Band
    4 kHz to 6 GHz
    Frequency Resolution
    1 Hz
    Frequency Error
    +/- 1 ppm per year
    Minimum Output Level
    -70 dBm
    Maximum Output Level
    +13 dBm (+7.0 dBm When Using AM)
    Amplitude Resolution
    0,1 dBm
    Amplitude Accuracy
    +/- 1 dB +/- 0,01 dB/dB
    Output Level Settling Time
    < 1 msec
    Harmonics
    <- 20 dBc
    Non-Harmonic Spurious
    <- 50 dBc
    Modulation Types
    CW, AM, FM, Pulse and Gated Pulse
    Modulation Frequency Range
    1 Hz to 100 kHz
    AM Modulation Depth
    5 to 95% (Usable 0 to 100%)
    AM Accuracy, Output LF
    Measured at 0 dBm Level
    < +/- 3% (5 to 95% Modulation Depth)
    AM Accuracy, Output HF
    Measured at 0 dBm Level
    < +/- 2% (10% to 90% Modulation Depth)
    < +/- 3% (5 to 10% and 90 to 95% Modulation Depth)
    Pulse On/Off Time Range
    200 ns to 100 s
    Pulse Modulation On/Off Ratio
    Output
    > 90 dB
    Supply
    12/28 VDC