Key Features

  • Conducts emissions measurements and EMI/EMC measurements
  • Ultra Broadband Frequency Range: 26 MHz to 40 GHz Radiated Immunity Compliance:
    • IEC 61000-4-3
  • Radiated Emissions Pre-Compliance:
    • Semi Anechoic (Pre-Scan)
    • ANSI C63.4/CISPR 16-1-4
    • Fully Anechoic
    • CISPR 16-1-4 (Fixed Height)
  • Turnkey Chamber Solutions

ETS-Lindgren's SpaceSaver FAR chambers are designed to perform IEC 61000-4-3 compliant radiated immunities measurements. These fully anechoic chambers can also be used for CISPR 16-1-4 pre-compliant radiated emissions measurement pre-scans, with fixed height. Frequency ranges for this chamber are 26 MHz to 40 GHz (radiated immunity) and 30 MHz to 40 GHz (radiated emission).

Specifications
View the SpaceSaver FAR Chamber technical specifications below

Physical Specifications

Scan Height: Fixed
Overall Structure Dim (H): 3.5 m (11.48 ft)
Overall Structure Dim (L): 7.8 m (25.59 ft)
Overall Structure Dim (W): 3.8 m (12.47 ft)
Internal Working Area (L): 7.6 m (24.93 ft)
Internal Working Area (W): 3.7 m (12.14 ft)
Internal Working Area (H): 3.4 m (11.15 ft)
Quiet Zone for Emission Testing (Diameter x H): 1.0 m x 0.5 m
(3.3 ft x 1.6 ft)
Field Uniformity: 0 to 6 dB; 1.5 m x 1.5 m (4.92 ft to 4.92 ft)

Electrical Specifications

Frequency Minimum: 26 MHz
Frequency Maximum: 40 GHz

Product Configuration

Frequency Minimum: 26 MHz
Frequency Maximum: 40 GHz