Key Product Information


​​​Fast Performance

The EMPower Pulse RF/Burst Power Meters perform power measurements at a maximum sampling speed of one million samples per second. By using such a high sampling rate, the power meter is able to measure RF burst/ pulse signals with a duration time as low as 2 μsec. Both power meters are also capable of measuring CW.

Accurate Measurements

EMPower Pulse allows high precision RF measurements with high dynamic range of 65 dB for the 6 GHz model, and 55 dB for the 18 GHz model.

Pulse-shape Recording Mode

In addition to the fast performance, the EMPower Pulse also has an internal measurement buffer that can store 4,000 samples. This enables the user to easily visualize RF burst signals.

Ford Radar Pulse Test

Fast speed and the record mode make the EMPower Pulse suitable for measurements in accordance to the Automotive Ford Standard EMC-CS-2009. 

Specifications Select a tab below for more information



Model
7002-003 (Sensor)​7002-005 (Sensor)
​​Detector Type​
​Diode
Diode​
Measurement Function​Peak Power​Peak Power​
​Calibrated Frequency Range
9 kHz to 6 GHz​80 MHz to 18 GHz​
​​Power Measuring Range​
​​-55 dBm to +10 dBm 
​​-45 dBm to +10 dBm​
​Input Damage Level 
​​>20 dBm​
​>20 dBm
​​Resolution
​ ​0.01 dB
​ ​0.01 dB
​​VSWR 
​​<1.05 @ 10 MHz to 100 MHz
<1.15 @ 100 MHz to 2 GHz
<1.35 @ 2 GHz to 6 GHz​ 
​<1.20 @ 80 MHz to 10 GHz 
<1.35 @ 10 GHz to 18 GHz
​​Maximum Linearity Error (0 dBm ref)​
​ 0.05 dB/10 dB
​ 0.05 dB/10 dB
​​Measuring Speed (Per Second)
​​1 MSps (10 MSps for CW)
100 kSps or 20 kSs (Software Selectable)​ 
​​​​1 MSps (10 MSps for CW)
100 kSps or 20 kSs (Software Selectable)​ 
​Logging Buffer Record Mode
4,000 Samples
2,000 Pre-trigger
2,000 Post-trigger​
​4,000 Samples
2,000 Pre-trigger
2,000 Post-trigger
​​Accuracy (23° ± 2° C) ​
​0.25 dB
​0.25 dB (≤10 GHz) 
0.50 dB (>10 GHz)​
​Temperature Effect ​
​<0.15 dB Over Full Temperature Range 
​<0.15 dB Over Full Temperature Range
​​ ​Measurement Units
​dBm 
​dBm 

Model
7002-001 (Modular Plug-in Card)​ 7002-003 (Sensor)​ 7002-005 (Sensor)​
​Temperature Range (Use)
​​0° C to +40° C
(32° F to 104° F)
​​​0° C to +40° C
(32° F to 104° F)
​​​0° C to +40° C
(32° F to 104° F)
​​Temperature Range (Storage)
​-20° C to +85° C
(4° F to 185° F)
​​-20° C to +85° C
(4° F to 185° F)
​​-20° C to +85° C
(4° F to 185° F)
​​Relative Humidity 
​​10 to 90% (Non-condensing)​
​​10 to 90% (Non-condensing)​
​​10 to 90% (Non-condensing)​
​Connector to Plug-in Card or PC (Data)
​n/a
​USB-B
​USB-B
​USB Communication
​n/a
​USB 1.1
​USB 1.1
​USB Power Consumption
​n/a
​< 200 mA​
​< 200 mA​​
​RF Input Connector ​
​n/a
​​N-type Precision
​​N-type Precision

  • EMPower RF Power Pulse Sensor (Please Specify Model)
  • 2 m Shielded USB Cable
  • Installation CD

 

 

EMCenter™ Modular RF Platform Model 7000-001 http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501001EMCenter™ Modular RF Platform Model 7000-001
EMPower™ Pulse RF/Burst Power Meter Model 7002-00X http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501005EMPower™ Pulse RF/Burst Power Meter Model 7002-00X
EMSwitch™ RF Switch Plug-in Cards 7001-00X http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501008EMSwitch™ RF Switch Plug-in Cards 7001-00X
EMControl™ Positioner Controller Plug-in Card Model 7006-01 http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501002EMControl™ Positioner Controller Plug-in Card Model 7006-01
EMPowerTM Power Meters Model 7002-00X http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501006EMPowerTM Power Meters Model 7002-00X
EMSense™ EMF Probe Plug-in Card Model 7007-002http://www.ets-lindgren.com/products/probes-monitors/emc-probes/electric-field-probes/5010/5010EMSense™ EMF Probe Plug-in Card Model 7007-002
EMPower™ Power Meters Model 7002-006 http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501005EMPower™ Power Meters Model 7002-006
EMGen™ RF Signal Generator Model 7003-001 http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501003EMGen™ RF Signal Generator Model 7003-001
EMSense™ EMF Probe Plug-in Card Model 7007-001 http://www.ets-lindgren.com/products/probes-monitors/emc-probes/electric-field-probes/5010/501007EMSense™ EMF Probe Plug-in Card Model 7007-001

Key Features


 

  • Fast Performance: Up to 1 Million Samples per Second
  • Accurate Measurements with Dynamic Range of:
    • -65 dB for 6 GHz Model (Model 7002-003)
    • -55 dB for 18 GHz Model (Model 7002-005)
  • Pulse-shaped Recording Mode Suitable for Ford Radar Pulse Test

 

 


Literature and Videos


 

 

http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501005<a href="/sites/etsauthor/ProductBrochures/General%20Brochures/Test_and_Measurement_Full_Line.pdf"><img alt="" src="/_layouts/15/IMAGES/icpdf.png" />Test_and_Measurement_Full_Line.pdf</a>

 

 

http://www.ets-lindgren.com/products/test-systems/rf-test-solutions/5010/501005

L o a d i n g

 Catalog-Item Reuse