Medical RF Filters 
​ETS-Lindgren's Medical RF Filters enable electrical power, lighting and data signals to be brought into the MRI room, through the RF shield. All RF filters are typically located in one area, usually near the penetration panel of the MRI equipment.

 

 


​​​ETS-Lindgren's Medical RF Filters enable electrical power, lighting and data signals to be brought into the MRI room, through the RF shield. All RF filters are typically located in one area, usually near the penetration panel of the MRI equipment. Below are the most common medical RF filters used, however ETS-Lindgren can provide virtually any filter you may need. UL, ULC and CE filters are also available, including low current leakage filters.

Facility filters allow specified frequencies to pass while suppressing or attenuating other frequencies. The filters listed as ELUL, EQ and LMF are our most common models utilized in medical applications. There are many other versions of RF filters also available for a wide variety of applications. Our complete line of RF filters incorporates a wide array of choices which provide effective EMI/RFI shielding to meet or exceed industry standards.

Key Product Information


Versatile Smaller Format

The LMF-4112 is a versatile filter that is utilized for standard lighting, switching and power circuits. This filter offers the advantage of an approximately 50% reduction in required surface area and weight. This means the use of more circuits with substantially less space and weight than standard filters offering the same features. This filter has a 0 - 30 amp rating and operates at 250 VAC/VDC. It is c-UL rated and CE marked. The EQ series are used for conventional thermostats, nurse calls, fire alarms, smoke detectors or any low leakage application requiring measurement of small variances in voltage to perform an application. This series of filters has an operating range of 1 to 10 amps and 24 VAC/VDC. It is also c-UL rated and CE marked. The ELUL filters are utilized for standard lighting, switching and power circuits. They are able to run at higher voltage requirements with a 0 - 30 amp range and operating up to 277 VAC and 480 VDC. It is c-UL rated and CE marked. 

Applications

ELUL, LMF and EQ filters are utilized to allow electrical services within a Magnetic Resonance Imaging suite. All applications that require a conductor within the room interface with a filter. The most common services are lighting, power, smoke detectors and fire alarms.


ELUL-2030 Medical RF Filter
EQ-2001-NR Medical RF Filter
 EQ-2010-1C Medical RF Filter LMF-3519 Medical RF Filter

 

Specifications Select a tab below for more information



​Model
ELUL-2030 
(Power, Light)​
EQ-2001-NR
(Smoke, Fire, Thermo.)​
EQ-2010-1C
(ERDU, EPO)​
LMF-3519
(CCTV)​
​Part Number
E-ELUL-2030​
​E-EQ-2001-NR
​E-EQ-2010-1C
​E-LMF-3519
​Rated Current (Amps)
0 to 30​0 to 1.0​0 to 10​0 to 10​
Max. Leakage Current
@220 V, 50 Hz
700 mA3 mA3 mA
N/A
Max. Leakage Current
@120 V, 60 Hz
450 mA1.5 mA
1.5 mAN/A
Rated Voltage0-277/480 VAC/VDC0-30/50 VAC/VDC0-250/250 VAC/VDC0-24 VDC
Insertion Loss100 dB min. from 150 kHz to 10 GHz per MIL-STD-220A.100 dB min. from 150 kHz to 18 GHz per MIL-STD-220A.100 dB min. from 150 kHz to 18 GHz per MIL-STD-220A.100 dB min from 30 MHz - 10 GHz per MIL-STD-220A
Voltage DropNot to exceed 1% of applied voltage at Unity Power Factor.Not to exceed 1% of applied voltage at Unity Power Factor.Not to exceed 1% of applied voltage at Unity Power Factor.N/A
SafetyDischarge resistors provided per UL-1283.Discharge resistors provided per UL-1283.Discharge resistors provided per UL-1283.N/A
Passband ImpedanceN/AN/AN/A75 (L-G)

​Model
ELUL-2030 
(Power, Light)​
EQ-2001-NR
(Smoke, Fire, Thermo.)​
EQ-2010-1C
(ERDU, EPO)​
LMF-3519
(CCTV)​
​Part Number
E-ELUL-2030​
​​E-EQ-2001-NR
​E-EQ-2010-1C
​​E-LMF-3519
Number of Lines
2
2
2
1​
Weight
6.8 kg (15.0 lb)
2.7 kg (6.0 lb)
2.7 kg (6.0 lb)

2.7 kg (6.0 lb)

Dimension A
53.98 cm (21.25 in)38.74 cm (15.25 in)
38.74 cm (15.25 in)38.74 cm (15.25 in)
Dimension B52.71 cm (20.75 in)37.47 cm (14.75 in)
37.47 cm (14.75 in)
37.47 cm (14.75 in)
Dimension C50.8 cm (20.0 in)35.56 cm (14.0 in)35.56 cm (14.0 in)35.56 cm (14.0 in)
Dimension D12.7 cm (5.0 in)7.62 cm (3.0 in).7.62 cm (3.0 in)7.62 cm (3.0 in)
Dimension E7.62 cm (3.0 in)5.08 cm (2.0 in)5.08 cm (2.0 in)
5.08 cm (2.0 in)
Dimension F3.96 cm (1.56 in)3.81 cm (1.5 in)3.81 cm (1.5 in)3.81 cm (1.5 in)


Dimension G
​7.62 cm (3.0 in)
5.08 cm (2.0 in)
5.08 cm (2.0 in)
5.08 cm (2.0 in)

  • Mounting Tabs Provide Secure Access
  • Feedthrough Penetration Pipe for Electrical Leads Functions as a Frequency Waveguide Cut-off
  • Single Phase, Two Line Filters
  • Seam Welded Case with Removable RF Gasketed Lid, Tin Coated for Long Wear

 

 

Medical RF Filters http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100813Medical RF Filters
Magnetic Active Compensation System (MACS) for MRI http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100802Magnetic Active Compensation System (MACS) for MRI
SafeScan® Ferromagnetic Detection Systems http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100810SafeScan® Ferromagnetic Detection Systems
Med-Vizion™ MRI Lighting LED Sheet Lamp Recessed Troffer LFP2x2-40Khttp://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100814Med-Vizion™ MRI Lighting LED Sheet Lamp Recessed Troffer LFP2x2-40K
Med-Vizion LED Down Light ZXR-3500 and ZXR-4000 http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100803Med-Vizion LED Down Light ZXR-3500 and ZXR-4000
SAFESCAN® Ferromagnetic Target Scanner™ http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100808SAFESCAN® Ferromagnetic Target Scanner™
Med-Vizion Graphic Display Panel (GDP) System http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100805Med-Vizion Graphic Display Panel (GDP) System
Med-Vizion™ MRI Lighting LED Sheet Lamp Recessed Troffer LFP2x4-40K http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100815Med-Vizion™ MRI Lighting LED Sheet Lamp Recessed Troffer LFP2x4-40K
OMS™ Sample Draw Oxygen Monitoring System http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100806OMS™ Sample Draw Oxygen Monitoring System
Automatic Opener/Closer For use with Auto-Seal II Doorhttp://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100801Automatic Opener/Closer For use with Auto-Seal II Door
CCTV Intercom System http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100804CCTV Intercom System
Safe IV Port for Medical Applications http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100809Safe IV Port for Medical Applications
EMI/RFI Shielded Waveguide Air Vents http://www.ets-lindgren.com/products/shielding/medical-shielding/medical-shielding-accessories/11008/1100811EMI/RFI Shielded Waveguide Air Vents

Key Features


  • Filter Models to Match Requirements for:
    • Power and Lighting
    • Low Leakage
  • 100dB Insertion Loss from 40 kHz
  • Voltage Drop Less than 1% of Applied Voltage
  • Compliant with:
    • Low Voltage Directive 2006/95/EC
    • MIL-STD-220C
  • UL 1283 Listed, 5th Edition