Home /
Categories /
Probes & Monitors /
HI-6153 Electric Field Probe
|
The laser powered HI-6153 Electric Field Probe provides broadband EMF frequency coverage and wide dynamic range that satisfies the demands of most test requirements. To take advantage of this capability, the HI-6153 was designed to be single range reading so data can be read continuously over the entire dynamic range. Data values for each axis (X,Y,Z) can be read individually, or summed.
|
Key Features
- Dynamic Range: 2.0 - 800 V/m (single range)
- Frequency Range: 10 MHz - 40 GHz
- Laser Powered for Unlimited Test Time
- Operates with ProbeView Laser Software or Most 3rd Party Immunity Packages
- Reads/Displays Individual Axis Summed Axis Values
- Suitable for Automotive Specs: SAE J1113/27, GM9120P
- Suitable for Commercial Specs: EN/IEC 61000-4-3 Radiated Immunities
- Suitable for Military Standards Specs: MIL-STD 461E Radiated Susceptibility (RS), DO-160
Description
The isotropic deviation (isotropy) of the HI-6153 is near ideal. This means the HI-6153 makes accurate field intensity measurements regardless of its orientation to the field of interest.
Fiber optic signal and power lines link the Model HI-6153 probe to either the model HI-6100 Field Monitor, or as a direct connect to a PC with the HI-6113 Laser Data Interface and ProbeView Laser Software.
The HI-6100 provides manual functions and programmed control via IEEE-48 and RS-232 Serial Data Interfaces. Readings from up to four probes can be displayed simultaneously with the HI-6100, and can be any combination of battery of laser powered probes.
The HI-6113 Laser Data Interface provides the laser power and communications for the HI-6153 Electric Field Probe. A USB port to the PC allows for quick and easy data collection, using ProbeView Laser software.
Standard Configuration
- 10m Fiber Optic Cable
- A2LA Standard Calibration
- Carrying Case
- Manual
- Probe Assembly
Electrical Specifications
|
Detection |
Isotropic (X, Y, and Z Axis Readings)
|
|
Frequency Range |
10 MHz to 40 GHz
|
|
Frequency Response (Nominal) |
10 MHz to 18 GHz +/-2.5 dB 18 GHz to 40 GHz +2.0, -4.0 dB
|
|
Dynamic Range |
2 to 800 V/m (Single Range)
|
|
Resolution |
0.01 V/m
|
|
Isotropic Deviation |
+/-1.0 dB
|
|
Linearity |
+/-0.5 dB of Reading
|
|
Sample Rate (Typical) |
HI-4413P Serial Modem - 44 samples/second maximum
HI-6100 Field Monitor - 70 samples/second maximum
|
|
Overload Withstand |
>1500 V/m CW
|
Physical Specifications
|
Mounting |
1/4-20 UNC Internal Thread
|
|
Physical Interface |
Duplex Optical Fiber (62.5 micron multimode) FC for laser, Integral 1m optical cable
|
|
Battery |
Rechargeable NiMH
|
|
Operating Range |
10° to 40° C, 5-95% Relative Humidity Non-condensing
|
|
Dim. L. |
43.8 cm
(17.24 inches)
|
|
Dim. W. |
5.7 cm
(2.24 inches)
|
|
Weight |
0.36 kg
(
0.79 lb )
|
|