Key Features
- A high performance, reliable concept in environmental shielding for MRI systems
- Can reduce fluctuations in excess of 120 mG to less than 1 mG
- Compatible with all magnets, from 0.2T to 3T and higher
- MACS negative feedback technology corrects itself even as the environment changes
- Offers protection against fluctuations in magnetic fields caused by: Subways, Transformers, Elevators and Moving Vehicles
- Uniform protection from environmental AC/DC magnetic interference over a wide frequency range
Description
While feed forward systems are less adept at changing with the environment, the MACS negative feedback technology corrects itself even as the environment changes.
All magnets are sensitive to abnormal fluctuations in environmental magnetic fields caused by moving vehicles, trains, elevators, electrical transformers, and other sources. Due to these interferences, many MRI site locations are rejected. Open, high field open, high field (1.5T) and super high field (3T and greater) can now be shielded with the MACS.
The MACS provides a uniform solution, at DC and through the low frequency spectrum, using a highly sophisticated electronic and magnetic compensation technology.
Applications
Providing real time compensation of changes in the environmental magnetic field, the MACS allows MRI systems to be placed in locations previously rejected due to the presence of varying high magnetic fields. By reducing these fluctuations experienced by the MRI system, image quality is maintained and defective scanning sequences are virtually eliminated.
Performance
Using a set of Helmholtz coils, a feedback loop and 3-axis sensing magnetometer installed in the room, the MACS monitors the external magnetic field environment and drives current through the coils to create an equal, but opposite, magnetic field. This produces a net effect of near zero magnetic field fluctuation experienced by the MRI system. Magnetic field fluctuations in excess of 120 mG can be reduced to less than 1 mG at magnet isocenter.
Maintenance-Free Operation
The MACS is designed to require no maintenance or adjustments following initial installation under normal operating conditions. Additionally, the system is designed for 24-hour operation.
The MACs reduces fluctuations
on the MRI system, allowing for higher
quality images (see before and after
images to the upper left).
The control/amplifier unit
(above)measures 20"x21"x8.5". System Installation
The MACS design requires just 2 to 3 days for installation and calibration with support of the magnet vendor. The system can be installed on new MRI sites, or on existing sites. The entire system is located inside the MRI suite, eliminating the need for additional filtering. The control/amplifier unit must be secured in the forward end of the room while the sensing probe is positioned in an optimum, yet unobtrusive location, usually behind the magnet. Coils are installed in a neat and inconspicuous manner, ensuring that the intended appearance of the MRI suite is not compromised.
Warranty
ETS-Lindgren's standard limited warranty covers parts and service for one year.